How to use this calculator:
- Enter the required thickness in microns. One micron is one thousandth of a millimetre.
- Enter the surface area you are plating in square centimetres (cm²).
- Select the type of swab you are using.
- The Calibration Factor defaults to 1 and can usually be left unchanged. See Note 1 below for calibration instructions.
- Click ‘Calculate’.
- Read the voltage setting and the volume of solution required.
- Use the Coulomb Counter reading (C) as your target charge. See Note 2 below.
- The final two results show the estimated weight of silver plated and the percentage of silver used from the calculated volume of solution.
- For work involving a substantial deposit of silver, greater accuracy can be achieved by using the Silver Pen Electrode rather than the Platinum Pen Electrode.
Click ‘Print Page’ to print a report containing your inputs and results.
Notes
1. Setting the calibration factor
When plating a number of identical items, you can calibrate the calculator by comparing the calculated deposit weight or thickness with the actual deposit.
Starting with a Calibration Factor of 1:
Calibration Factor = calculated weight or thickness ÷ actual weight or thickness.
For example, the calculator predicts a deposit weight of 0.32 grams at a thickness of 2.5 microns. If the measured weight gain is 0.35 grams, the Calibration Factor is:
0.32 ÷ 0.35 = 0.91.
If you are recalibrating after using a factor other than 1, multiply the existing Calibration Factor by the calculated-to-actual ratio.
The weight method requires suitably accurate scales. Weigh the article clean and dry before and after plating.
If you have access to X-ray fluorescence or microsection measurements, you can use the measured thickness to establish the Calibration Factor. Further information is available in the Process Sequences tab on the Coulometer product page.
2. Using the coulometer
The Coulomb Counter reading (C) is the target charge for the job. Start from zero and plate until the counter reaches this value, using the selected swab and the voltage shown by the calculator.
Avoid creating a short circuit: charge passed through a short circuit does not contribute to the deposit and can make the reading inaccurate or invalidate the result.